5300 Series Manual Dual-Bank Manifold Systems
5300 Series Manual Dual-Bank Manifold Systems
5300 series manual dual-bank manifold system consists of a main gas delivery bank and a reserve bank of cylinders. When the primary cylinder bank is depleted, manually “turn off” the valve on the primary bank and open the valve on the reserve bank to reactivate gas flow. The changeover of this system needs to be operated manually.
Features
Open-style manifold
Silver brazing on piping joints for maximum leak prevention
System is designed to accomodate future expansion needs
System is mounted with gas filters
Pressure switch port is available
Headers have been tested to withstand high cylinder pressure
Wall or floor mount available
Installation Dimensions
Installation Dimensions
Manifold System Layouts
Manifold System Layouts
Ordering Information
Gas Service Series Max. Inlet Pressure Delivery Pressure Max. Delivery Flow Outlet Connection Pigtail Specifications
Oxygen
5300X
3000 psi (207 bar)
50~125 psi (3.5~8.6 bar)
3500 SCFH (100 m3/h)
3/4" NPT (M)
Pigtail, CGA540
Acetylene
5300Y
435 psi (30 bar)
2~15 psi (0.14~1 bar)
700 SCFH (20 m3/h)
3/4" NPT (M)
Pigtail, CGA510
Propane
5300F
435 psi (30 bar)
15~125 psi (1~8.6 bar)
1050 SCFH (30 m3/h)
3/4" NPT (M)
Pigtail, CGA510
Carbon Dioxide
5300C
3000 psi (207 bar)
50~125 psi (3.5~8.6 bar)
2100 SCFH (60 m3/h)
3/4" NPT (M)
Pigtail, CGA320
Argon, Nitrogen, Helium
5300IN
3000 psi (207 bar)
50~125 psi (3.5~8.6 bar)
3500 SCFH (100 m3/h)
3/4" NPT (M)
Pigtail, CGA580
Air
5300Q
3000 psi (207 bar)
50~125 psi (3.5~8.6 bar)
3500 SCFH (100 m3/h)
3/4" NPT (M)
Pigtail, CGA346
Hydrogen
5300H
3000 psi (207 bar)
50~125 psi (3.5~8.6 bar)
10500 SCFH (300 m3/h)
3/4" NPT (M)
Pigtail, CGA350
EX: 53
Series
1
Manifold System Layout
2
Cylinder Valve Spacing
X
Gas Service
- 5 x 5
Number of Cylinders (left-hand / right-hand)
- 1
Type of Mounting
53 1: Standard layout
2: “L” Shape layout
3: “U” shape layout
4: Crossover layout
5: Staggered layout
1: 5” (127 mm)
2: 10” (254 mm)
3: 13” (330 mm)
4: 18” (457 mm)
X: Oxygen
Y: Acetylene
F: Propane
C: Carbon Dioxide
IN: Ar, He, N2
Q: Air
H: Hydrogen
1 x 2: One cylinder on the left,
Two cylinders on the Right
5 x 5: Five cylinders on the left,
Five cylinders on the Right
...
Note: Direction of piping (Right or Left) is
indicated by facing the manifold.
1: Wall mount
2: Floor mount

Example: 5312X-5x5-1 indicates a 5 x 5 cylinder dual-bank manifold system.

Distance between two cylinders is 10” on standard horizontal layout.